Ultraviolet germicidal irradiation (UVGI) systems are gaining popularity, however objective comparisons of their characteristics are lacking. While environmental cultures and reduction of hospital-associated infections rates are excellent study endpoints, they are impractical for centers with limited resources who want to compare or optimize UVGI systems use.
Methods
We evaluated radiometry and commercial test cards, two simple and low cost tools, to compare 2 full size UVGI systems (Tru-D and Optimum-UV Enlight) and 2 small units (Lumalier EDU 435 and MRSA-UV Turbo-UV).
Results
Radiometry-derived output curves show that if both large devices emit enough energy to reach C. difficile lethal doses at 10 ft, the reduction in output in distance is almost perfectly logarithmic. In a patient room environment, Enlight and Tru-D performed similarly when compared using radiometry and commercial test cards. The two small devices reached C. difficile range around the bathroom with the device raised above the floor, but longer times are needed.
Conclusions
Despite different workflows and price points, no clear superiority emerges between Tru-D and Enlight. Bathroom disinfection should be dealt with separately from the main room and small, cheaper units can be used. Radiometry and commercial test cards are promising ways to compare UVGI systems, but further validation is needed using correlation with environmental cultures.
The Full Report: Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards